6) To secure the coaxial plug, insert the retaining ring
into the groove of the lamination ring using tweez
ers.
7) Then install the electric wrist rotator in the lamina
tion ring.
6.13 Wiring without electric wrist rotator
1) One of the two three-pole connectors is marked
with a coloured dot (arrow). The two receptacles
must not be interchanged!
INFORMATION: The two-pole motor cable must
not be attached to the coaxial plug.
2) Install the plastic screw to secure the two connect
ors.
3) Push the coaxial plug in to the stop and rotate it
until the stop noticeably engages.
4) Install the coaxial plug in the lamination ring.
7 Handling
7.1 Charging the battery
The product is powered by an integrated high-quality Li-Ion battery with sufficient capacity to operate for one full
day of usual everyday activities. Turning the product off during longer periods of passive use (e.g. air or rail travel,
visit to a theatre or cinema, etc.) will make the battery last longer. We recommend charging the product once a day
when used by the patient on a daily basis. Electronic battery management provides the patient with information
about the battery charge level (battery management).
The charging unit consists of the charging plug and the power cord. The battery charger has an input voltage
range of 100–240 V and may be operated in a mains frequency range of 50–60 Hz.
INFORMATION
Please note the corresponding instructions for use of the battery charger.
53
DynamicArm Plus 12K110N=*
Handling
Summary of Contents for DynamicArm Plus
Page 71: ...71 DynamicArm Plus 12K110N...