We recommend charging the product once a day when used by the patient on a daily basis.
•
The battery should be charged for at least 3 hours prior to initial use.
•
Note the permissible temperature range for charging the battery (see page 68).
7.2 Data transfer between the product and the PC
Product settings using the adjustment software can only be made via Bluetooth data transfer. For this purpose, a
Bluetooth wireless connection must be established between the product and the PC using the "60X5 BionicLink
PC" Bluetooth adapter. The installation and use of the "60X5 BionicLink PC" adapter are described in the instruc
tions for use supplied with the adapter.
7.3 “ElbowSoft TMR” software
INFORMATION
For the installation/removal of ElbowSoft TMR*, observe the 647G1230 instructions for use.
7.3.1 Calibration
Calibration establishes the patient’s movement range (maximum desired extension, maximum desired flexion). A
movement beyond these two established thresholds using myosignals is not possible.
Heavy clothing can reduce the movement range. Nevertheless the elbow joint attempts to achieve the maximum
movement range and maximum flexion angle through myosignals.
INFORMATION
Calibration takes place in the course of the initial fitting, after adapting a new or modified socket, after each repair
and after every service.
The elbow joint cannot be adjusted if the calibration process has not been completed yet.
1) Hold the forearm of the product.
2) Start the calibration process by clicking the “Start”
button on the “Calibration” tab.
INFORMATION: The elbow lock is released
after clicking the “Start” button. Therefore, sup
port the forearm of the product before calibra
tion starts.
3) With the free hand, move the product to the desired
maximum extension.
4) With the free hand, move the product to the desired
maximum flexion (15° to 145°).
INFORMATION: The effective flexion angle is
determined on the patient by the shape and
size of the prosthetic socket and by clothing.
5) Click the “End” button to complete the calibration
process.
55
DynamicArm Plus 12K110N=*
Handling