43
Charging recept
acle
Beeper
Explanation/correction
-
3 x short
Axon-Bus prosthetic system switches itself
off
•
Charge battery
-
Pulsating
Opening the Axon-Bus gripping component
in an emergency
•
System is switched off, system can be used
without restriction after being switched on
again
Battery charger
Explanation/correction
An error has occurred in the Axon-Bus prosthetic sys
tem
•
Consult prosthetist
Axon-Bus prosthetic system requires service
•
Consult prosthetist
Error while charging the product
Battery charger
Explanation/correction
Axon-Bus prosthetic system is not charging
•
Disconnect and then reconnect the charging plug.
•
Clean charger contacts.
Charger is overheated or overcooled
•
Disconnect charging plug and allow charger to cool
down/warm up.
Battery charger or power supply unit is defective
•
Consult prosthetist
12.2.3 Status signals
Battery charger is connected
Battery charger
Explanation
Power supply and battery charger operational
Battery is charging
Battery is charging, battery 25% charged
Summary of Contents for Axon-Bus
Page 2: ...1 2 3 4 5 6 2...
Page 262: ...262 3 2 mm...
Page 263: ...263 3 3 3 mm 0 12 8E600 2 1 1 2 3 3 3 1 3 2 1 274 3 3...
Page 264: ...264 3 4 3 5 Ottobock 4 4 1 4 2 1 1 2 4 3...
Page 265: ...265 AC AC 4 4...
Page 266: ...266 CT MRI WiFi 30 cm 274 4 5...
Page 267: ...267 Axon Bus Axon Bus Axon Bus 453H10 1 Ottobock 453H10 1 Ottobock 453H16 Ottobock...
Page 268: ...268 4 6 4 7 12K501 12K500...
Page 270: ...270 6 4 7 7 1 1 1 7 2 1 3 2 7 3 1 2...
Page 271: ...271 1 1 2 4 3 7 4 1 5 6 2 1 5 6 2 9S501 9S503...
Page 272: ...272...
Page 277: ...277 PPPP YYYY WW PPPP YYYY WW 12 2 12 2 1 1 2 1 2 1 12 2 2 1 3...
Page 278: ...278 12 2 3 AC 25 50 75 100 100 50...
Page 282: ...282...
Page 283: ......