14 | Ottobock
6 MyoWrist Transcarpal 10V38=* for the fitting of short
residual limbs
Combination with a 8R1=* Endoskeletal-Adapter is necessary for this fit-
ting (Fig. 1).
Before finishing a short socket, please be aware of the limited amount of
space inside the socket.
6.1 Determining socket length
Attach the 8E44=* Transcarpal Hand to the MyoWrist* Transcarpal 10V38=*,
but do not secure the nuts with Loctite. (Fig. 3).
Next, pre-assemble the tube of the Endoskeletal-Adapter to the MyoWrist
Transcarpal 10V38=*, without bonding the tube in place. For further pro-
ceeding, please refer to the Instructions Manual of the Endoskeletal-Adapter
647H501.
6.2 Assembly instructions
After having cut the tube to the desired length and having assembled the
Endoskeletal-Adapter, please proceed as follows (Fig. 4):
Clean all threads (except the two bolts in the plate) with acetone. Do not
touch the cleaned areas! After the acetone has evaporated, coat the fol-
lowing threads with 636K12 Loctite 245: Threaded Connector with Tube,
Threaded Connector. Coat the Chassis with Quick-Disconnect Wrist with
636K13 Loctite 241.
Feed the four-pole cable through the tube adapter until it comes out the
other end. Screw the Chassis with Quick-Disconnect Wrist together with the
threaded connector (hand tight, approx. 8 Nm). Use pliers with plastic jaws
for this purpose. Feed the four-pole cable through the MyoWrist Transcarpal
10V38=* and tighten (approx. 8 Nm) to the Endoskeletal-Adapter by hand.
Please be aware, that the cable can become twisted in this process
(Fig. 5)
. The cable must straightly pass through the wrist unit in
order not to be damaged (Fig. 6).
Slide the Cable Seal over the Four-Pole Cable (Fig. 7) and press the Cable
Seal into the plate of the MyoWrist Transcarpal (Fig. 8).