PRODUCT SETTINGS AND CONFIGURATION
32
OI-6000-X-X-X-X-O OPS_GUIDE_REV 3.1
3.9
CALIBRATION METHOD
The calibration method selection allows you to choose how you calibrate the sensor element. Manual calibration is
the default method for all gas types.
Manual calibration lets you use the ADD and SUB buttons during calibration to match the reading shown
on the screen to the value of the gas being applied.
Auto calibration will set the reading, after a predetermined amount of time, during calibration to the value
entered during the auto calibration setup process.
1.
Use the ADD button to select manual calibration and the SUB button to select auto calibration.
2.
Press the MENU button to advance to the Modbus Address Setting screen.