DSPEC Pro
®
Digital Gamma-Ray Spectrometer User’s Manual
794380D / 0914
56
CLEAR_ALL
This command is equivalent to the combination of CLEAR_COUNTER, CLEAR_DATA,
CLEAR_PRESETS, and CLEAR_ROI commands.
CLEAR_COUNTER
The live time and true time counters are set to zero.
CLEAR_DATA
The channels of spectral data in the window of interest (see SET_WINDOW command) are
set to zero. The ROI flags are not changed, nor are the presets changed.
CLEAR_HV_HEALTH
Clears (resets) the state-of-health bits. This command clears the SOH bits in the DIM and
does not affect the stored SOH (which is reported when the instrument is not acquiring data).
CLEAR_LIST
Clears data and resets acquisition in List Mode. The response is the same as the CLEAR
command.
CLEAR_MDA_PRESET
Clears all MCA preset values. This command is used to quickly clear the list of MDA preset
values.
CLEAR_PRESETS
The live time, true time, ROI integral, ROI peak, and overflow presets are all set to zero
(disabled).
CLEAR_ROI
The region-of-interest flags for the channels in the window of interest (see SET_WINDOW
command) in the currently selected device (see SET_DEVICE command) are cleared.
CLOSE_FILE_FLASH crc
No function in the DSPEC Pro; included for backward compatibility.
COMP
No function in the DSPEC Pro; included for backward compatibility.
DISABLE_BLRE_AUTOMATIC
Disables the automatic determination of baseline restorer time constant. See also ENABLE_
BLRE_AUTOMATIC, SET_BLRE, and SHOW_BLRE.