DSPEC Pro
®
Digital Gamma-Ray Spectrometer User’s Manual
794380D / 0914
86
SHOW_TIME_ETP
Returns the current Enhanced Throughput Mode protection time.
Response:
TIME_ETP 0000000000012.8
Protection time is set to 12.8
µ
s.
SHOW_TIME_START
Reports the time of the last START command in the form
$Nhhhmmmsssccc<CR>
where hhh represents a 3-digit integer hour (0 through 23), mmm represents a 3-digit integer
minute (0 through 59) and sss represents a 3-digit integer second (0 through 59). See also
SET_TIME_START, SET_DATE_START, and SHOW_DATE_START.
Example Responses:
$N010054017052<CR>
Time returned 10:54:17 (10 h, 54 min, 17 s).
$N020013037050<CR>
Time returned 20:13:37 (20 h, 13 min, 37 s).
SHOW_TRIGGER
Tells whether the trigger is enabled or disabled. See also ENABLE_TRIGGER.
Responses:
$IF
Trigger disabled.
$IT
Trigger enabled.
SHOW_TRIGGER_POLARITY
This returns the polarity setting of the input trigger for the sample changer (1 means the
trigger is positive-edged and triggers on a low-to-high transition; 0 means it is negative-edge
triggered and triggers on a high-to-low transition). See also TRIGGER, ENABLE_
TRIGGER, DISABLE_TRIGGER, SHOW_TRIGGER, and SET_TRIGGER_POLARITY.
Responses:
$FNEG<CR>
The HV is set to negative.
$FPOS<CR>
The HV is set to positive.
SHOW_TRIGGER_SAMPLE
Reports the trigger source in Insight Mode.
Responses:
$F0LLD
LLD is source.
$F1PKD
Peak-detect is source.
$F2RANDOM
Trigger happens randomly.
$F3Gate
Gate Input triggers the waveform.
SHOW_TRIGGER_SPECIAL
Not used; included for backward compatibility.
Summary of Contents for DSPEC Pro