794380D / 0914
4. FIRMWARE COMMANDS AND RESPONSES
69
Bit 1:
Negative baseline threshold.
Bit 2:
Baseline restorer gate.
Bit 3:
Positive baseline threshold.
Bit 4:
Busy.
Bit 5:
Gate.
Bit 6:
Reserved.
Bit 7:
Peak detect.
Bit 8:
Reserved.
SET_ROI_UNCERTAINTY start, chans
Sets the region to be used for the uncertainty preset calculation. See also SHOW_ROI_
UNCERTAINTY.
SET_SEGMENT
Not used; included for backward compatibility.
SET_SHAP_FLAT value
Sets the width of the flattop to value. Value is in microseconds, ranging from 0.3 to 2.4
µ
s in
steps of 0.1. See also SHOW_SHAP_FLAT.
SET_SHAP_RISE value
Sets the rise time to value. Value is in microseconds, ranging from 0.8 to 23
µ
s in steps of
0.2. See also SHOW_SHAP_RISE.
SET_SHUTDOWN index
Used in conjunction with the LIST_SHUTDOWN command to set the high voltage
shutdown setting. Index specifies a position in the list of legal shutdown settings which were
displayed with the LIST_SHUTDOWN command.
Example:
SET_SHUT 0
Sets the SHUTDOWN to TTL.
SET_SHUT 1
Sets the SHUTDOWN to ORTEC.
SET_SHUT 2
Sets the SHUTDOWN to SMART.
SET_SHUTDOWN_ORTEC
This sets the bias shutdown mode to ORTEC.
SET_SHUTDOWN_SMART
Sets the bias shutdown mode to SMART.
SET_SHUTDOWN_TTL
This sets the bias shutdown mode to TTL.