DSPEC Pro
®
Digital Gamma-Ray Spectrometer User’s Manual
794380D / 0914
82
SHOW_PEAK_CHANNEL
This command returns the number of the ROI channel with the largest number of counts. An
ROI channel is a channel that has the ROI flag set. The lowest number ROI channel with the
largest count is reported if more that one channel contains the largest number of counts.
Channel 16383 is the highest numbered channel in any device.
Responses:
$C00000087<CR>
Maximum count was found in channel 0 or no ROI channels
were found.
$C00001088<CR>
Maximum count was found in channel 1.
...
...
$C16382107<CR>
Maximum count was found in channel 16382.
$C16383108<CR>
Maximum count was found in channel 16383.
SHOW_PEAK_PRESET
Reports the value of the ROI peak preset. See SET_PEAK_PRESET for information about
the ROI peak preset.
Responses:
$G0000000000075<CR>
Peak preset disabled.
$G0000000001076<CR>
Peak preset set to 1 count.
$G2147483646120<CR>
Peak preset set to 2147483646 counts.
$G2147483647121<CR>
Peak preset set to 2147483647 counts.
SHOW_PZ
Displays the PZ setting. See also SET_PZ.
Responses:
PZ_ENA 002200
Pole zero enabled and set to 2200.
PZ_DIS 001000
Pole zero disabled and set to 1000.
SHOW_PZ_AUTOMATIC
Shows the state of the automatic PZ. See also SET_PZ_AUTOMATIC.
Responses:
$IT<CR>
Pole zero enabled.
$IF<CR>
Pole zero disabled (TRP).
SHOW_RADIX
SHOW_RATE_ZDT
SHOW_REMOTE
Not used; included for backward compatibility.
SHOW_RENH
Reports if Resolution Enhancer mode is enabled.