794380D / 0914
4. FIRMWARE COMMANDS AND RESPONSES
61
INITIALIZE_ZERO_STABILIZATION
Resets the zero peak stabilization adjustment to unity (no adjustment). This value is reported
as 2048 by the SHOW_ZERO_ADJUSTMENT command. See also SET_ZERO_
ADJUSTMENT, ENABLE_ZERO_STABILIZATION, and DISABLE_ZERO_
STABILIZATION.
LIST_BLRE
Lists the valid settings of the baseline restore function. See also SET_BLRE.
Response:
BLRE 10.0 100.0 10 100
Legal settings are 10.0 to 100.0 in steps
of 1.
LIST_CORR_FLAT
Lists the valid settings of the flattop width. See also SET_CORR_FLAT.
Response:
CORR_FLAT
!
1.0 .992 0 255
Legal settings are from
!
1 to 0.992 in
(255
!
0 + 1 = 256) steps.
LIST_DISPLAY
Not used by the DSPEC Pro; included for backward compatibility.
LIST_GAIN_COARSE
Lists the valid coarse-gain settings.
Response:
GAIN_COAR 1 2 4 8 16 32
LIST_GAIN_CONV
Lists legal conversion gain settings.
Response:
GAIN_CONV 512 1024 2048 4096 8192 16384
LIST_GAIN_FINE
Lists the valid fine-gain settings.
Response:
GAIN_FINE 0.45 1.0 88437 196607
0.45 to 1 in 108171 steps.
LIST_HV
Lists the valid high-voltage settings.
Response:
HV 0 5000 0 5000
0 to 5000 in 5000 steps.