91
Figure 26. The SOH Status Tab within MAESTRO.
APPENDIX A. STATE-OF-HEALTH BIT
DEFINITIONS
Following are descriptions of the SOH bit. To monitor these within MAESTRO, click Acquire/
MCB Properties..., then on the Status tab (see Fig. 26).
Bit
Description
Comment
0
+24V
The first four bits monitor the voltages coming from the DSPEC
Pro. If one of these values varies by more than ±2 V, the corre-
sponding bit is set.
1
+12V
2
-12 V
3
-24V
4
Temperature
Set if the detector temperature increases and shuts down the HV
(therefore, the SHUTDOWN bit will also be set).
5
HV voltage
Set if the high voltage deviates by more than ±300 V from the
value you entered in MAESTRO.
6
Overload
Set if the preamplifier overloads for more than about 4.0 seconds.
7
Shutdown
Set if the preamplifier shuts down for any reason.