4. FIRMWARE COMMANDS AND RESPONSES
63
OFF
Turns the DSPEC Pro off.
OPEN_FILE “filename.ext”
OPEN_FILE_FLASH
RESET
RESET_REMOTE
No function in the DSPEC Pro; included for backward compatibility.
SET_BLRE baseline
This sets the baseline restorer time constant to the value baseline. Baseline is in micro-
seconds with a range from 10 to 100. The specified value is only used if automatic baseline
restorer mode is disabled. See also ENABLE_BLRE, DISABLE_BLRE, and SHOW_BLRE.
SET_CORRECTION_FLAT value
This sets the flattop correction to value. The flattop correction is normally determined with
the OPTIMIZE command. See also SHOW_CORRECTION_FLAT.
SET_DATA [start, chans],value
If the optional start and chans parameters are included in this command, the range of chan-
nels specified by start and chans is loaded with value. If start and chans are not specified,
sets all channels of spectral data in the window of interest (see SET_WINDOW command)
to the specified value. ROI flags are not affected.
SET_DATA_APPLICATION “string1",”string2"
This is used to store information in the DSPEC Pro internal memory that can be used by
other programs, such as sample descriptions and energy calibrations. C
ONNECTIONS
uses this
feature. String1 = the data identifier, 32 bytes maximum; string2 = the data, 128 bytes
maximum.
SET_DATE day,month,year
Sets the date stored in the battery backed-up system clock to the specified values. Day can be
any value from 1–31; month any value from 1–12; and year any value from 0–99. The cur-
rent date and time are stored for a device when an acquisition is started. See also SHOW_
DATE, SET_TIME, and SHOW_TIME.
SET_DATE_START day,month,year
Sets the start date to the specified values. Normally the start date and time are set automati-
cally whenever a device is started with the START command. See also SHOW_DATE_
START, SET_TIME_START, and SHOW_TIME_START.