Figure 7-35: The acquired image is too far away
the images are marked with the message “
Image not suitable for CT processing
”
(that is
image not suitable for topography
) as indicated by the arrows in Figure 7-36
Figure 7-36: Images are marked as
as not being suitable for corneal topography processing
In these cases, to prevent artifacts, it is advisable to discard the images using
[Discard] and repeat the acquisition.
As noted above, after the topo-aberrometric images have been acquired, it is
possible to process them locally or transmit them to the external PC for processing
and/or filing following the test.