6.3
INSTALLATION PROCEDURE
After the Keratron
™
Onda has been placed on the support surface and the wheel
protection covers have been placed in their position, the connections are to be
made to the power supply and, if necessary, to a personal computer.
As shown in the figure, remove the plastic cover located on the front of the
instrument.
Figure 6-1: Removing the front cover
Starting from the left side it is possible to use the label (Figure 6-2) to identify the
socket for the LAN connection, the socket for the power supply cable, and a socket
for the USB connection or optional Wi-Fi key.
Figure 6-2: Connector housing label
The cables are inserted from beneath the table through the appropriate hole in the
support surface itself and attached in such a manner as to give them a sufficient
amount of play to avoid both obstruction and pulling during all of the movements of
the X-Y-Z cart.
Connect to the power outlet only the power cable provided or a cable with
equivalent safety characteristics.
An external personal computer can be connected directly to the Keratron
™ Onda
using the Ethernet LAN “
crossed
” cable provided, or an equivalent cable (see §9.4).