HiScan Touch
Installation and Operation Manual
OPTIKON 2000
Code 14101XEN
5-11
2014-09-02 Rev.B
5.3.2 GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC IMMUNITY
HiScan is intended to be used in the electromagnetic environment specified below.
The customer or user of HiScan should assure that it is used in such environment.
Immunity test
EN 60601 test level
Compliance level
Electromagnetic Environment – Guidance
Electrostatic discharge
(ESD)
EN 61000-4-2
•
±6 kV contact
•
±8 kV air
•
±6 kV contact
•
±8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical/fast
transient/burst
EN 61000-4-4
•
±2 kV for power supply lines
•
±1 kV for input/output lines
•
±2 kV for power supply lines
•
±1 kV for input/output lines
Mains power quality should be that of a typical
commercial or hospital environment
Surge
EN 61000-4-5
•
±1 kV differential mode
•
± 2kV common mode
•
±1 kV differential mode
•
± 2kV common mode
Mains power quality should be that of a typical
commercial or hospital environment
Voltage dips, short
interruptions, and
voltage variations on
power supply input lines
EN 61000-4-11
•
<5% U
T
(>95% dip in U
T
) for
0.5 cycles
•
40% U
T
(60% dip in U
T
) for 5
cycles
•
70% U
T
(30% dip in U
T
) for
25 cycles
•
<5% U
T
(>95% dip in U
T
) for
5 seconds
•
<5% U
T
(>95% dip in U
T
) for
0.5 cycles
•
40% U
T
(60% dip in U
T
) for 5
cycles
•
70% U
T
(30% dip in U
T
) for
25 cycles
•
<5% U
T
(>95% dip in U
T
) for
5 seconds
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of HiScan requires continued operation during
power mains interruptions, it is recommended
that the Keratron® Scout be powered from an
uninterruptible power supply or a battery.
Power frequency (50/60
Hz) magnetic field
EN 61000-4-8
3 A/m
Not applicable
The device does not contain any
magnetically sensitive
components.
Power frequency magnetic fields should be at
level characteristic of a typical location in a
typical commercial or hospital environment.
Conducted RF
EN 61000-4-6
Radiated RF
EN 61000-4-3
3 Vrms
From 150 kHz to 80 MHz
3V/m
From 80 MHz to 2.5 GHz
3 Vrms
3V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of HiScan, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d= 1.2
√
P
d= 1.2
√
P From 80MHz to 800 MHz
d= 2.3
√
P From 800MHz to 2.5 GHz
Where
P
is the maximum output power rating to
the transmitter in watts (W) according to the
transmitter manufacturer and
d
is the
recommended separation distance in metres
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbo
l: