HiScan Touch
Installation and Operation Manual
OPTIKON 2000
Code 14101XEN
3-3
2014-09-02 Rev.B
•
In case of doubt on the size, perform a calibration test on the probe. If the test
fails or the probe is not working correctly, do not use the device and contact the
technical support.
Pachymetry
•
Gently apply the probe on the moistened cornea, being careful not to alter its
normal geometry.
•
In case of doubt on the size, perform a calibration test on the probe. If the test
fails or the probe is not working correctly, do not use the device and contact the
technical support.
Echography
•
Gently apply the probe on the patient's eye using ultrasound gel or another
substance suitable to facilitate the penetration of the ultrasounds in the tissues.
•
When using an immersion probe, use the special “mini-bath” cup with sclera
support.
•
The immersion cup must always be filled with sterile solution.
•
Keep the cup in place with one hand without applying an excessive pressure. If
liquid leaks from the sclera supporting ring, apply a thin layer of methocel to
enhance adherence.
•
Remove the liquid from the cup using a syringe without needle.
•
Clean and sterilise the probes before using them on another patient or applying
them to the other eye of the patient.
Footswitch
•
Never lift or move footswitch by the cable to avoid possible damages.
•
The footswitch is not autoclaveable.
•
The pedal can be used to activate the following functions:
o
Freeze/Unfreeze – Starts/stops Bscan, Biometry, Pachymetry or AScan
scans
o
Save – Saves the current scan
o
Rec – Starts/stops the recording of a movie