HiScan Touch
Installation and Operation Manual
OPTIKON 2000
Code 14101XEN
5-10
2014-09-02 Rev.B
5.3
ELECTROMAGNETIC COMPATIBILITY TABLES
5.3.1
GUIDANCE AND MANUFACTURER’S DECLARATION –
ELECTROMAGNETIC EMISSIONS
HiScan is intended to be used in the electromagnetic environment specified below.
The customer or user of HiScan should assure that it is used in such environment.
Emissions test
Compliance
Electromagnetic Environment –
Guidance
RF emissions
CISPR 11
Group 1
HiScan uses RF energy for its
internal functions only. Therefore, its
RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
HiScan is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low voltage
supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
EN 61000-3-2
Class A
Voltage
fluctuations/Flicker
emissions
EN 61000-3-3
Complies