HiScan Touch
Installation and Operation Manual
OPTIKON 2000
Code 14101XEN
9-2
2014-09-02 Rev.B
instructions provided with the accessories. A disposable protective cap should
always be used in presence of UBM.
9.4 MAINTENANCE
9.4.1 DEVICE
Store the echograph in a clean dry location at room temperature.
NOTE
To keep the device safe, check the device's conformity with electric current leakage limits according to
standard EN60601-1 at least once a year. Contact the bioengineering division of your facility or your
Optikon 2000 authorised distributor/service centre.
9.4.2 ACCESSORIES
1) Avoid dropping or mishandling probes and accessories. It is critical that these
components be handled with outmost care and inspected thoroughly after each
use for any damage or wear.
2) Place the probes on the support after each use. Avoid placing the probes on a
table: a potential fall could damage the tip and make it unusable.
3) Cover the probes with the protective caps if you are not planning to use them for
long periods of time.