Note 1: At 80 MHz and 800 MHz, the separation distance for the highest frequency range applies.
Note 2: These guidelines are not applicable in all situations. Electromagnetic propagation is influenced by the absorption and reflection of structures, objects and people.
a
The intensities fiels for stationary transmitters, such as the base stations for radio telephones (cellular and cordless) and terrestrial radiomobiles, radioamateur devices,
transmitters in AM/FM and TV transmitters can not be foreseen theoretically with precision. An electromagnetic survey of the site should be considered to evaluate an
electromagnetic environment caused by fixed RF transmitters. If the field strength measured in the place do to the device is used exceeds the applicable level of
compliance above, the operation of the device should be observed. If you notice abnormal performance, additional measures may be required as a different orientation
or position of the device
b
The field strength in the frequency range from 150 khz to 80 MHz should be less than 3 V/M
Recommended distances separation between portable and mobile radiocommunication devices.
The device is intended to operate in a environment electromagnetic space where RF radiated disturbances are under control. The customer or the user of the device can help
prevent electromagnetic interference by ensuring a minimum distance between mobile and portable RF communication devices (transmitters) and the device as recommended,
in relation to the maximum output power of the radio equipment.
Maximum output power of the
specified transmitter
W
Separation distance for transmitter frequency
m
From 150 kHz a to MHz
d = 1,17 √P
from 80 MHz to 800 MHz
d = 1,17 √P
from 800 MHz to 2,5 GHz
d = 2,34 √P
0,01
0,12
0,12
0,24
0,1
0,37
0,37
0,74
1
1,17
1,17
2,34
10
3,70
3,70
7,40
100
11,70
11,70
23,40
For the transmitters specified for maximum output power not shown above, the recommended separation distance d in meters (m) can be calculated using the equation
applicable to the frequency of the transmitter, where p is the maximum rated output power of the transmitter in Watts (W) according to the manufacturer of the transmitter
NOTE 1: to 80 MHz and 800 MHz, applies the separation distance for the highest frequency range.
NOTE 2: These guidelines may not be applied in all situations. Electromagnetic propagation is influenced by the absorption and reflection of structures, objects and people.
Summary of Contents for LINEA ESSE
Page 1: ...Instruction handbook Dental Unit LINEA ESSE...
Page 5: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 5 Version with floor spittoon unit...
Page 50: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 50 Note Check for condensate on a weekly basis...
Page 69: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 69...
Page 70: ......