OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
9
DANGER
The power switch isolates the equipment from the mains electricity. So, before performing any
operations in the equipment, make sure that the power switch is off.
WARNING
The instructions found in this manual and in the wiring diagram must be followed to connect a
suction unit; the suction unit must have EC markings as per directive 93/42/EEC as amended
“Medical Equipment” and meet international safety regulations IEC EN 60601-1 (Medical
electrical equipment – General safety regulations), IEC EN 60601-1-2 (Medical electrical
equipment – Collateral standard: Electromagnetic compatibility).
DANGER
In the version with the spittoon unit on the floor, when the chair moves upwards, check the
position of the patient's arm on the armrest, to prevent the arm from hitting the basin.
DANGER
When moving the chair, both in manual and especially in automatic mode, do not place hands
and feet near the bottom of the chair to avoid crushing hazards, read the chair instructions
carefully.
WARNING
During automatic chair movement, the operator must stand near the dental unit.