OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
21
Doing the same operation again makes you revert to the starting situation.
Spray is pre-selecting entering the command
on the main keyboard. Also see paragraph
MICROMOTOR OPERATION.
The water delivered to the spray is regulated by opening/closing the tap located under the micromotor
module: when turned anticlockwise, the flow of water to the spray is gradually reduced; when turned
clockwise, the opposite occurs. We recommend not using other regulation devices located alongside the spray
tap.
If the system is equipped with an optic fibre lighting system, the lights go on when the control
installed in the main table keyboard.
Note:
For further details refer to the manufacturer’s operating instructions enclosed.
SCALER MODULE
The tooth scaler operates when the instrument is lifted from its housing and moved towards the operating
field, and when the foot control is engaged (see paragraph TOOTH SCALER OPERATION).
When the instrument is lifted from the table, all the other dynamic instruments and the patient chair
movements are disabled.
Pressing control
to pre-select the spray. See the SCALER OPERATION paragraph.
The water delivered to the spray is regulated by opening/closing the tap located under the tooth scaler
module: when turned anticlockwise, the flow of water to the spray is gradually reduced; when turned
clockwise, the opposite applies.
Some scalers are fitted with the Endo and Perio modes.
EMS scaler:
-
to use the Endo function, the appropriate tip must be fitted and the necessary power (between 10
and 30%) must be set on the display. Attention, in any case 30% of maximum power must never be
exceeded in Endo mode.
-
to use the Perio function, the appropriate tip must be fitted and the necessary power (between 10
and 50%) must be set on the display. Attention, in any case 50% of maximum power must never be
exceeded in Perio mode.
The regulation of the power of the SATELEC SP NEWTRON scaler has to be done according to the tip on the
handpiece.
Summary of Contents for LINEA ESSE
Page 1: ...Instruction handbook Dental Unit LINEA ESSE...
Page 5: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 5 Version with floor spittoon unit...
Page 50: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 50 Note Check for condensate on a weekly basis...
Page 69: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 69...
Page 70: ......