OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
16
3.
COMMAND - CONTROL - WARNING
SAFETY SYSTEMS
The unit is equipped with safety devices that minimize the risk of collision during the movement of the chair, in
particular there is a safety on the basin floor water group in version, which intervenes when the tray is rotated
by the side of the chair, limiting the ascent of the chair at a defined height.
DANGER
In the version with a floor water group during the ascent of the chair check the
positioning patient on the arm to prevent the risk of crushing with the basin.
Note:
During any automatic chair movement (reset, memory placement or rinse position) by
pressing any patient chair command (from the dental unit or push off of the foot controller)
the chair stops immediately.
The movement of the chair, either at power-up that in normal use, can be prevented if one or
more safety systems are active
Note:
Lifting the assistant's table locks the pantograph and backrest downward movement.