OMS LINEA ESSE S.T.01/3S ED.0 REV.14 03/2017
41
CLEANING AND DISINFECTION
Activity
Frequency
By whom?
Note/references
Instrument table, instrument tube
exterior
After each patient
Trained
operator
Only use products indicated by OMS.
Instrument table:
instrument holder, handles, tray mat
Assistant table mat
After each patient
Trained
operator
Autoclave (if applicable)
Only use products indicated by OMS.
Instruments and syringe
After each patient
Trained
operator
Clean, disinfect and sterilise when
indicated by the instrument
manufacturer’s instructions.
Chair: Upholstery, headrest and
armrests
After each patient
Trained
operator
Only use products indicated by OMS.
Operating lamp components
After each patient
Trained
operator
Clean, disinfect and sterilise when
indicated by the manufacturer’s
instructions.
Foot control cleaning
Daily
Trained
operator
Only use products indicated by OMS.
Spittoon, cannula and spittoon filter
Daily
Trained
operator
Only use products indicated by OMS.
INSTRUMENTS
For instrument cleaning, sterilization and lubrication, we recommend consulting the instructions supplied by
the manufacturer(s) for each model that are attached to this manual and added to the equipment packaging.
SPITTON
The ceramic surface of the spittoon allows for rapid and easy daily cleaning using suitable products. The
spittoon output water cannula can easily be removed for cleaning and sterilisation in an autoclave, if
necessary.
Summary of Contents for LINEA ESSE
Page 1: ...Instruction handbook Dental Unit LINEA ESSE...
Page 5: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 5 Version with floor spittoon unit...
Page 50: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 50 Note Check for condensate on a weekly basis...
Page 69: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 69...
Page 70: ......