Version with floor spittoon unit
1.
Remove the equipment from its packaging;
2.
if required, disassemble the chair lifting the cover positioned between the column and the base,
disconnecting the electrical cables and removing the screws securing the two bases together;
3.
position the column in the previously prepared area following the installation plan;
4.
adjust the feet under the base and use a level to make the surface on the patient chair support flange
perfectly horizontal;
5.
re-assemble the patient chair and turn the adjustment feet correctly until possible “gaps” are
registered, screw the chair to the floor with the two expansion bolts supplied;
6.
reconnect the cables and fit back the cover that had previously been removed;
7.
install the patient chair seat and secure it (from underneath) with the 4 screws supplied;
8.
remove the spittoon unit guard using the supplied wrench and make the required connections;
9.
install the microscope arm and make the electrical connection;
10.
fit the monitor (if any) and connect it electrically together with the relative supply unit;
11.
if an operating lamp is featured, install it and connect it to the mains;
12.
check and, if necessary, connect the grounding cables to the screw identified by the grounding system
symbol;
13.
for the connection of the suction pump (see the wiring diagram), follow the same instructions as for
the power supply cables described in the next paragraph (cable fixing and traction).
14.
connect the water and air systems, followed by the drain and electrical systems. The electrical
network must be connected only to the appropriate terminal block in the floor box, while all the other
connections can be done either in the floor box or inside the spittoon unit.
Summary of Contents for LINEA ESSE
Page 1: ...Instruction handbook Dental Unit LINEA ESSE...
Page 5: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 5 Version with floor spittoon unit...
Page 50: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 50 Note Check for condensate on a weekly basis...
Page 69: ...OMS LINEA ESSE S T 01 3S ED 0 REV 14 03 2017 69...
Page 70: ......