83
4. Press the Up Key to display
yes
and press the Mode Key.
Set Value LED Display Model
Basic Model
5. The warning message
sure
will be displayed if field calibration is performed
for the first time. OMRON does not guarantee the results of field calibration.
This warning message will not be displayed the next time field calibration is
performed on the K3NX.
Set Value LED Display Model
Basic Model
6. Press the Mode Key to display the
inĆt
input range setting.
Set Value LED Display Model
Basic Model
7. Press the Shift Key so that the K3NX will be ready for the input range.
Set Value LED Display Model
Basic Model
8. Press the Up Key and input the range to be calibrated.
Set Value LED Display Model
Basic Model
9. The teaching indicator will be lit and
l.ref
will be displayed for reference in-
put signal L.
Set Value LED Display Model
Basic Model
10. Press the RESET/TEACH Key and input 1.0000 V as reference input signal
L. The teaching indicator will flash.
Set Value LED Display Model
Basic Model
Maintenance Mode
Section 5-3