34
Power Supply Frequency
•
Set the number of sampling times in the setup menu according to the power
supply frequency in order to shut off inductive noise that may be generated
from the power supply.
Setting
Unit
Number of sampling
times
Default
50/60
Hz
50: 12.5 times/second
60: 15 times/second
50
Follow the steps described below to set the power supply frequency to 60 Hz.
Set Value LED Display Model
Basic Model
1, 2, 3...
1. Press the Mode Key for more than one second while the
setup
setup menu
is displayed. The
inĆt
input range setting will appear.
Set Value LED Display Model
Basic Model
2. Press the Mode Key to display the
fre
power supply frequency setting.
Set Value LED Display Model
Basic Model
3. Press the Shift Key to display the prior set value
50
for changing. The PROG
indicator will flash.
Set Value LED Display Model
Basic Model
4. Press the Up Key to display the set value
60
. The displayed value will be
validated automatically if no change is made for five seconds. The
fre
pow-
er supply frequency setting will be displayed again.
Set Value LED Display Model
Basic Model
fre
FUNCTION
SETTING
SETTING
EXAMPLE
Setting Mode
Section 3-2
AUDIN - 7 bis rue de Tinqueux - 51100 Reims - France - Tel : 03.26.04.20.21 - Fax : 03.26.04.28.20 - Web : http: www.audin.fr - Email : [email protected]