Movement Average Processing
When an input is sampled, the new sampling data is added to the previous sam-
pling data and averaged.
Movement average processing is ideal for removing cyclic noise from the input
signal.
The number of sampling times can be set to 2, 8, 16, or 32.
Follow the steps described below to set the number of movement average processing
times to 16.
Set Value LED Display Model
Basic Model
1, 2, 3...
1. Press the Mode Key for more than one second while the
opt
option menu is
displayed. The
aUg
average processing setting will appear.
Set Value LED Display Model
Basic Model
2. Press the Shift Key to display the prior setting
1
for changing. The PROG
indicator will flash.
Set Value LED Display Model
Basic Model
3. Repeatedly press the Up Key until
m0016
is displayed. The setting will be
validated automatically if no change is made for five seconds. The
aUg
aver-
age processing setting will be displayed again.
Note Press the Mode Key to enter the set value immediately. The next pa-
rameter will be displayed for setting.
Set Value LED Display Model
Basic Model
When no operation is executed for five seconds
Set Value LED Display Model
Basic Model
SETTING
EXAMPLE
Setting Mode
Section 3-2