82
5-3-4 Reference Input Signal
The following are the reference signals for all input ranges.
Input type
Input range
Reference
input signal L
Reference
input signal H
Input
terminals
DC voltage
a Ud
0.00 V
199.99 V
1 and 4
input
b Ud
0.000 V
19.999 V
2 and 4
c Ud
0.0000 V
1.9999 V
3 and 4
d Ud
0.00 mV
199.99 mV
3 and 4
e Ud
1.0000 V
5.0000 V
2 and 4
DC current
a ad
0.00 mA
199.99 mA
1 and 4
input
b ad
0.000 mA
19.999 mA
2 and 4
c ad
0.0000 mA
1.9999 mA
3 and 4
d ad
4.000 mA
20.000 mA
2 and 4
AC voltage
a Ua
0.0 V
400.0 V
1 and 4
input
b Ua
0.00 V
199.99 V
1 and 4
c Ua
0.000 V
19.999 V
2 and 4
d Ua
0.0000 V
1.9999 A
3 and 4
AC current
a aa
0.000 A
10.000 A
1 and 4
input
b aa
0.0000 A
1.9999 A
1 and 4
c aa
0.00 mA
199.99 mA
2 and 4
d aa
0.000 mA
19.999 mA
3 and 4
Follow the steps described below to calibrate the input range
eĆUd
.
Set Value LED Display Model
Basic Model
1, 2, 3...
1. Press the Mode Key for more than one second while the
mant
maintenance
mode setting is displayed. The
reĆin
initialization setting will be displayed.
Set Value LED Display Model
Basic Model
2. Press the Mode Key to display the
calib
field calibration setting.
Set Value LED Display Model
Basic Model
3. Press the Shift Key so that the K3NX will be ready for the field calibration.
Set Value LED Display Model
Basic Model
SETTING
EXAMPLE
Maintenance Mode
Section 5-3