4. Setting (Air Flow Station Operation)
4-16
4.2.6 Changing Air Flow Sensor Set Value
This section explains the procedure to change Air Flow Sensor set values.
Follow the steps below.
(1) Display the ID of the target Air Flow Sensor unit in RUN mode.
(Refer to "5.3 Screen Transition in RUN Mode" for details of the operation procedure.)
Display
(Upper/Lower)
Operation
01.F01
Integrated Standard
Flow Rate
("RUN" Blinking)
Press the MODE key repeatedly until "RUN" starts blinking.
The Air Flow Sensor unit ID and measured value code is displayed
at the upper row.
The measured value is displayed at the lower row. The integrated
standard flow rate is displayed here.
A long press of the
△
key switches Air Flow Sensor unit IDs. Hold
the key until the ID of the target unit appears.
↓
Hold
△
Key
Display
(Upper/Lower)
Operation
03.F01
Integrated Standard
Flow Rate
For example, display "03" as the ID.
(2) Switch the mode to FUN mode and set NN.SET to "DISP".
Display
(Upper/Lower)
Operation
CYCLE
1s
("FUN" Blinking)
Press the MODE key repeatedly until "FUN" starts blinking.
(The status indicator of the ID03 Air Flow Sensor blinks in yellow
when the unit enters FUN mode.)
"CYCLE" is displayed at the upper row.
Press the
▽
or
△
key to display NN.SET.
↓
▽
or
△
Key
Display
(Upper/Lower)
Operation
NN.SET
OFF
Press the SET/REC/STOP key. The set value at the lower row
starts blinking.
↓
SET/REC/STOP Key
Display
(Upper/Lower)
Operation
NN.SET
OFF
↑Blinking
Press the
▽
or
△
key to display DISP, and then, press the
SET/REC/STOP key.
↓
SET/REC/STOP Key
Display
(Upper/Lower)
Operation
NN.SET
DISP
↑Blinking
The setting is applied when DISP stops blinking.