2. Part Name and Function
2-3
2.1.3 Input/Output Specifications
(1)
Breakout Cable Wires Color Code (D6FZ-JD
A)
The following shows the specifications of the breakout cable connected to the cable
connector socket of an Air Flow Sensor unit.
(2)
Analog Output
Analog Output 1 and Analog Output 2 are available for single Air Flow Sensor operation.
The specifications are shown below:
Output method
Source current (4-20 mA) method
Output accuracy
±
0.1 mA (Except for measurement accuracy)
Load resistance
Max. 270 Ω
Output detail
(*)
Analog Output 1 Standard Flow Rate (L/min (std))
Full-Scale Flow Rate: 1000 L/min (std)( *)
0 L/min(std): 4 mA
1000 L/min(std): 20 mA ( *)
Analog Output 2 Pressure (kPa)
Full-Scale: 1000kPa (1MPa)
0Pa: 4 mA
1Mpa: 20 mA
(*) Can be changed by the D6FZ-FGX21 Air Flow Station.