OMNI 4000/7000 Operations and Maintenance Guide
– Rev F
3-22
OMNICONNECT® Operations
To set up the plenum pressure option, follow these instructions:
1.
In the Actions ribbon, click
Configure
.
2.
a. Expand the
Prover
function in
the
Configure
tree.
b. Select
Configuration and
Setup
.
c. Verify that a
Pulse Interpolation
type prover is selected in the
Prover Type
screen.
3.
a. Click on
General Setup
in the
Configure
tree.
b. Select the
Equipment List
tab in
the screens on the right.
4.
a. In the
Prover
group on the lower
right-hand side of the screen,
check the
Plenum Pressure
option.
b. Click
Apply
to save your change.