OMNI 4000/7000 Operations and Maintenance Guide
– Rev F
7-25
Maintenance
1.
a. From the
Home
screen on the front
panel display, use the arrow keys to
navigate to and select
Utilities
.
b. Press the
OK
key.
2.
a. On the
Utilities
screen, use the arrow
keys to navigate to the
Calibration
menu selection.
b. Press
OK
.
3.
With
Input Calibration
highlighted, press
OK
.
The
Process Input Calibration
screen includes a horizontal Input bar at the top with
these options:
•
Process Input (PI) allows you to calibrate by channel number selection
(for example, 1 through 32).
•
Numbers 1 through 6 allow you to calibrate using a process input variable name
for individual Meters 1 through 6 (for example, Meter 1 Temperature or Meter 5
Density Pressure).
•
S1 and S2 allow you to calibrate using a process input variable name for
Individual Station 1 and Station 2 (for example, Station 1 Density Temperature or
Station 2 Density Pressure).
4.
Use the arrow navigation keys to select an
option on the
Input
bar; then press
OK.