55
Appendix e
iLog error messages
Error # Description
Note
-8003
User stopped logging readings.
-10005 Failed to find the Receiver.
Ethernet cable is disconnected,
Receiver is powered off,
connections across the firewall
require longer “connection to socket
time out” setting.
-10006 Windows socket was closed.
-10007 Windows socket error.
Wrong IP or wrong Port number
was used.
-10008 The Receiver failed to respond to a request.
Wrong IP or wrong Port number
was used.
-10011 Response came empty.
No data was sent.
-10012 Device responded with
Possibly the iLog is configured for
"Serial Time Out" string.
wrong product model.
-10014 Terminal Server Mode when the Port is 1000.
Try Port 2000 in iLog configuration.
-15100 Error on obtaining the temperature reading.
Possibly the iLog is configured for
wrong product model.