App-2
IM
Makes the status byte mask setting/queries the current
setting.
Syntax for setting
IMm <terminator>
m=0 to 63
1: Measurement end
2: Output change end
4: Syntax error
8: Overrange
16: Loop power supply error
32: Error on output
Syntax for query
IM? <terminator>
Example of response
IM6
Description
:
• Set the value representing the
information to be retrieved. If
more than one bit is to be
retrieved, the sum of the number
of those bits must be set as "m."
For instance, if 3 bits,
"Measurement end," "Syntax
error," and "Overrange," are to
be retrieved, set "m" to 13
(1+4+8=13).
• For details on status bytes, see
page App-5.
MF
Sets the measuring function/queries the current setting.
Syntax for setting
MFm <terminator>
m=0: DCV
1: DCA
2: Resistance
Syntax for query
MF? <terminator>
Example of response
MF1
MO
Starts/Stops measurement/queries the current setting.
Syntax for setting
MOm <terminator>
m=0: Stop
1: Start
Syntax for query
MO? <terminator>
Example of response
MO1
MR
Sets the measurement range/queries the current setting.
Syntax for setting
MRm <terminator>
DCV
DCA
Resistance
m= 0: 500 mV 20 mA 500
Ω
1:5V
100 mA 5 k
Ω
2:35 V
-
50 k
Ω
Syntax for query
MR? <terminator>
Example of response
MR0
OB
Queries the battery charge state.
Syntax for query
OB <terminator>
Description
The following are the responses:
0: Off
1: Charging
2: Charging completed
OD
Outputs measured data.
Syntax for query
OD <terminator>
Description
For the output format of the
measured data, see page App-5.
OE
Outputs error information.
Syntax for query
OE <terminator>
Example of response
ERR11
Description
• Outputs the most-recently-
generated error.
• For the meanings of error
numbers, see the troubleshooting
section (page 40).
OR
Queries whether an RJC sensor is connected.
Syntax for query
OR <terminator>
Description
The following are the responses:
0: Connected
1: Not connected
OS
Outputs the setting information.
Syntax for query
OS <terminator>
Description
• For the output format of the
measured data, see page App-5.
SC
Enable/Disable auto-zero calibration/queries the current
setting.
Syntax for setting
SCm <terminator>
m=0: Enable auto-zero calibration
1: Disable auto-zero calibration
Syntax for query
SC? <terminator>
Example of response
SC0