Check Before Calibrating the Resistance Generation Function
Generate the one of the appropriate resistance values listed in the following table from
the instrument, and measure the output voltage using a DMM. If the measured value
deviates from the tolerance, calibration is unavailable when generating resistance.
Contact your nearest representative.
Range
Generated Resistance
Tolerance
500
Ω
0.00
Ω
±
20
µ
V or less
5 k
Ω
0.0000 k
Ω
±
20
µ
V or less
50 k
Ω
0.000 k
Ω
±
200
µ
V or less
Wiring
SOURCE
Hi
Hi
mA
120mA MAX
ALL TERMINALS 42V
PEAK MAX TO GND.
250V F 125mA
FUSE
MAX
22mA
42V MAX
28V
Lo
Lo
MEASURE
I+
I-
Guard
Hi
Lo
Ω
Guard
Calibrated instrument
Measure DCV.
CAUTION
• Do not apply a voltage exceeding the maximum input voltage;
otherwise, the input part may be damaged.
• Do not short-circuit or apply an external voltage to output
terminals of the instrument or standard equipment, or else their
internal circuitry may be damaged.
Calibrating the Generation Function
Wiring
DCV
SOURCE
Hi
Hi
mA
120mA MAX
ALL TERMINALS 42V
PEAK MAX TO GND.
250V F 125mA
FUSE
MAX
22mA
42V MAX
28V
Lo
Lo
MEASURE
I+
I-
Guard
Hi
Lo
Ω
Guard
Calibrated instrument
Measure DCV.
DCA
SOURCE
Hi
Hi
mA
120mA MAX
ALL TERMINALS 42V
PEAK MAX TO GND.
250V F 125mA
FUSE
MAX
22mA
42V MAX
28V
Lo
Lo
MEASURE
I+
I-
Guard
Hi
Lo
Ω
Guard
Calibrated instrument
Measure DCA.
Free Datasheet http://www.datasheet-pdf.com/