![background image](http://html2.mh-extra.com/html/omega/compact-cal-ca100/compact-cal-ca100_owners-manual_5154776012.webp)
Components and Their Functions
11
7
Output on and off key
Turns output on and off.
8
+/- key
Toggles the polarity of the output value.
9
NEXT ENTER key
This switches the associated settings when setting the output values for the
frequency and pulse signal. It also fixes the entered value when using the numeric
keypad (refer to the following description).
10 ZERO CLR key
This resets the output value set to zero when using the up/down key (refer to the
following description). When using the numeric keypad (refer to the following
description), the value being entered is canceled and the previous set value is
restored.
11 Output value setting keys
This sets the output value for the generation function. Either of the following two
key modes can be selected from the menu. The up/down key mode is selected as the
factory default setting.
Up/down key: This increments or decrements the values by one count for each digit
corresponding to the [ ]/[ ]. If you try to increment or decrement using this key
with value 9 or 0, the current digit moves up or down by one digit.
Numeric keypad: This enters numbers 0 to 9 directly from the keypad.
12 Decimal point key
This enters the decimal point when using the numeric keypad (refer to the above
description). When this key is pressed, the digits to the left of the decimal point
move to the positions specified on a range basis. When using the up/down key, this
key is unavailable.
Operation keys for measurement
13 Display hold key
This retains displayed values as they are.
14 Range selection key
This selects the measurement range.
15 Function selection key
This selects one of the measuring functions: voltage, current, or resistance.
16 Measurement on and off key
This turns measurement on and off.
LCD Screen
DCV
Ω
Ω
Ω
HOLD
MEASURE
mV
mV
mA
mA
kHz
k C
k
SOURCE
AVERAGE
CHARGE END
TC
DCA
DCV
DCA
RTD
FRQ
PULSE
ON
OFF
INT RJC BRKE JTN
Pt100
24V OUT
SOURCE
ITS-90 IPTS-68
MEASURE
1
2
3
4
5
6
8
10
11
9
7
Ω
CYCLES
F
1
State-of-charge indicator
When using Ni-Cd batteries (optional), this indicates its charge status (CHARGE
indicates charging is taking place and CHARGE END indicates charging has
completed).
2
Low-battery indicator
This lights up when the battery becomes weak.