26
Hi
Lo
Lo
mA
Hi
Lo
C
R
CA100
SOURCE
(DC mA)
(99020)
MEASURE
(DC mA)
Input source:
Equipment such as a
converter
R: 200
Ω
±
10%, 1/4 W
C: 1
µ
F
±
10%, 50 V
(Equivalent to the 533M5002105K resistor from Matsuo Electric)
Note however that this additional circuitry reevaluates the CA100’s response specification as noted
below.
Response: 1 sec (at load resistances no greater than 2 k
Ω
)
Do not use this circuitry for purposes other than current generation; otherwise, it can produce
measurement errors.
Thermocouple, Resistance Temperature Detector
This generates thermo-electric power corresponding to the temperature of thermocouple
(TC) or resistance corresponding to the temperature of resistance temperature detector
(RTD) from the output terminal.
1
Press [FUNCTION] on the SOURCE side and align to [TC] or [RTD].
2
Press [RANGE] on the SOURCE side to select TC type.
Type
TC/RTD Type
Display
Temperature Range
TC
K
E
J
T
N
B
R
˚C
˚C
˚C
˚C
˚C
˚C
˚C
-200.0 to +1372.0˚C
-250.0 to +1000.0˚C
-210.0 to +1200.0˚C
-250.0 to +400.0˚C
-200.0 to +1300.0˚C
+400 to +1820˚C
-40 to +1767˚C
RTD
Pt100
˚C
-200 to +850˚C
* RTD : The specifications are compatible with both IEC 751-1983 and IEC 751-1995.
TC
: The specifications are compatible with both IEC 584-1-1989 and IEC 584-1-1995.
To switch the standard, refer to page 21.
3
To change the polarity, press the [+/-] key.
Selecting "-" places a leading minus sign (-) before figures, while selecting "+"
places no leading sign.
4
Specify the temperature.
• Using [ ] or [ ] (for initial setting)
Press [ ] or [ ] to specify the output value from the rightmost digit.
To reset the value to zero, press the [ZERO CLR] key.
• Using the numeric keypad (refer to page 21)
Enter a temperature value using the numeric keypad and the decimal point key
(where applicable), and press the [NEXT ENTER] key. The figure flickers when
entering a value.
When entering decimals, press the decimal point key first and then the
appropriate key or keys on the numeric keypad.
If a value out of the generation range is specified, an error occurs. In this case,
press [ZERO CLR] to reset to the previous value, and reenter a new value.
5
Press the [SOURCE ON] key to start generation.
[ON] is displayed.
To stop the output, press [SOURCE ON] again.
[OFF] is displayed.