Before Starting Generation or Measurement
21
Note
When backlighting the instrument, the internal power consumption increases, and the internal
temperature rises. This may cause a degraded accuracy in generation or measurement compared with
a normal condition. For accuracy, refer to the section, "Specifications," on page 51.
Averaging, Key Type, International Temperature, Temperature Unit Settings
Settings
This instrument allows the setting of Averaging, Key Type, International Temperature,
and Temperature Unit Settings from the maintenance menu.
Averaging setting:
This specifies whether to enable (on) or disable (off) the moving
average of the measured data. If the measured data display
fluctuates due to noise, set the averaging setting to on to perform
the moving averages. The setting defaults to off.
Key-type selection: Either the up/down keys (UP-DN) or numeric keypad (TEN) is
available for the output value setting key. The setting defaults to
"UP-DN."
International temperature standard selection: Either of IPTS68 or ITS90 is available for
the international temperature standard. The setting defaults to
"ITS90."
Selection of temperature unit: Either ˚C or ˚F can be selected as the temperature unit.
The default setting is ˚C.
Setting
1
Press the [NEXT ENTER] key and the [ZERO CLR] key at the same time.
[SP FC] and [End] are displayed.
2
Press the [ ] or [ ] key until [End] on the lower line of the display changes to [Set
uP], and then press the [NEXT ENTER] key.
The setup menu is displayed.
3
Pressing the [ ] or [ ] key changes the bottom menu from [AVG] to [KEY] and [t
tYPE]. To set the averaging function, display [AVG] and press the [NEXT ENTER]
key.
4
Select [on] or [oFF] using [ ] or [ ] and press the [NEXT ENTER] key.
5
To specify the key type, display [KEY] using [ ] or [ ] and press the [NEXT
ENTER] key.
6
Select [uP-dn] or [tEn] using [ ] or [ ] and press the [NEXT ENTER] key.
7
To set the international temperature standard, display [t tYPE] using [ ] or [ ] and
press the [NEXT ENTER] key.
8
Select the [iPtS68] or [itS90] using the [ ] or [ ] and press the [NEXT ENTER]
key.
9
To set the temperature unit, display [t unit] using the [ ] or [ ] and press the
[NEXT ENTER] key.
10
Select [C] or [F] using the [ ] or [ ] and press the [NEXT ENTER] key.