6.1 Reprocessing
SC-1 INSTRUCTION MANUAL
Ch.6
6
Prepare another lint-free cloth moistened with
disinfectant solution (medical-grade 70% ethyl
alcohol, 70% isopropyl alcohol or not greater
0.65%(6500ppm) sodium hypochlorite).
Figure 6.3
7
Disinfect the surfaces that previously contained patient debris (refer to Step 5) by
wiping with the moistened lint-free cloth.
8
Ensure that the surfaces are completely wet for the contact time instructed by the
disinfectant solution manufacture.
9
If any surface of this instrument remains wet, wipe it with a dry lint-free cloth and let it
dry thoroughly.
Lint-free cloth moistened with
disinfectant solution