6.1 Reprocessing
67
SC-1 INSTRUCTION MANUAL
Ch.6
Preparing equipment for reprocessing
Equipment needed
The following equipment is prepared to perform the reprocessing steps described in this chapter.
Personal protective equipment (examples)
Eyewear
Face mask
Moisture-resistant
clothing
Chemical-resistant
gloves
*1
*1 Long sleeve gloves are recommended to prevent skin exposure.
Chemicals used for reprocessing
• Detergent solution and disinfectant solution
(Refer to “
Detergent solution and disinfectant solution”)
Other
• Clean lint-free cloths
*2
*2 Lint-free cloths are recommended for reprocessing to prevent lint or cloth fibers from lodging or being trapped
in the instrument’s components. If gauze is used to reprocess the container, ensure that fibers do not get
caught on.