6.1 Reprocessing
69
SC-1 INSTRUCTION MANUAL
Ch.6
Disinfection
WARNING
Follow the usage (temperature, contact time, use life) and amount (concentration)
provided by detergent solution manufacturer and disinfectant solution
manufacturer.
Failure to follow manufacturer's instructions may result in insufficient cleaning and
disinfection effect.
CAUTION
Wipe remaining detergent solution and disinfectant solution according to
instructions of detergent solution manufacturer or disinfectant solution
manufacturer. Failure to do so may adversely affect the human body or this
instrument.
1
Shut down the tablet following Section 5.5, “Power off”.
2
Disconnect the AC adapter from the cradle.
3
Undock the tablet from the cradle.
4
Prepare a lint-free cloth moistened with neutral
detergent solution
Figure 6.2
5
Wipe off all patient debris from this instrument using the moistened lint-free cloth.
Make sure that you wipe LCD lightly so LCD will not be damaged.
Lint-free cloth moistened with
detergent solution
Summary of Contents for SC1
Page 2: ......
Page 6: ...iv Contents SC 1 INSTRUCTION MANUAL ...
Page 18: ...12 Important Information Please Read Before Use SC 1 INSTRUCTION MANUAL ...
Page 40: ...34 3 6 Creating an account SC 1 INSTRUCTION MANUAL Ch 3 8 Tap Next Figure 3 12 Next ...
Page 64: ...58 4 7 Sleep mode SC 1 INSTRUCTION MANUAL Ch 4 ...
Page 70: ...64 5 5 Power off SC 1 INSTRUCTION MANUAL Ch 5 ...
Page 80: ...74 6 3 Disposal SC 1 INSTRUCTION MANUAL Ch 6 ...
Page 88: ...82 8 3 Returning the tablet for repair SC 1 INSTRUCTION MANUAL Ch 8 ...
Page 94: ...App 88 Specifications SC 1 INSTRUCTION MANUAL ...