51
MX63/MX63L
51
1
Setting the analyzer (when using U-MDICAF3)
When using the mirror unit (U-MDIC3), it is not necessary to set
the analyzer or adjust the crossed nicols.
If the DIC slider is engaged in the light path, remove it from the light
path. For details, see “Setting the DIC slider" (P.52).
Engage the 10X objective or 20X objective in the light path and bring
the sample into focus approximately.
If the cover or the dummy slider is attached to the analyzer insertion
slot a , remove it.
Insert the analyzer (U-AN360-3) b in the analyzer insertion slot a .
Analyzer insertion position
Light path
First level (pulled out)
Empty hole
Second level (pressed in)
Analyzer
Rotate the analyzer rotation dial c to adjust the analyzer.
Observation with crossed nicols*:
Set the index of the dial to the
position shown in the picture on the
right.
Observation without crossed nicols*:
Rotate the dial while looking at the observed image and align it with
the position where the desired image is visible.
* The crossed nicols means the state that the field of view
becomes the darkest.
During DIC observation, the analyzer rotation dial must be set
to the crossed nicols state.
TIP
1
2
3
4
55
Dial
Index
TIP
b
c
a