Loosen the mounting knob a at the rear right of the revolving
nosepiece and pull out the dummy slider b .
Insert the MIX slider for reflected light observation (U-MIXR) c to the
slider insertion slot of the revolving nosepiece so that the connector
comes to the right side, and push it in the second level (position where
the clicking sound is heard).
MIX slider for reflected light
observation position
Light path
First level (pulled out)
OUT
Second level (pressed in)
IN
You can check the state whether the MIX slider for reflected light
observation is engaged in or removed from the light path by the
POSITION indicator d of the hand switch (BX3M-HS).
Tighten the mounting knob a to secure the slider.
1
2
3
d
a
1
c
b
2
1
Inserting the MIX slider for reflected light
observation
a
b
c
If you press the ON/OFF button a of the hand switch (BX3M-HS) to set
it to ON, the illumination of the MIX slider for reflected light observation
(U-MIXR) turns ON.
State
Indicator ( b , c )
Function
ON
Turns ON.
The illumination turns ON.
OFF
Turns OFF.
The illumination turns OFF.
1
2
Turning ON the illumination of the MIX slider
for reflected light observation