111
MX63/MX63L
111
3
Attaching the reflected light filter (26ND0.5)
Loosen the clamping screws on the top cover a using the Allen screwdriver (2 mm) to remove the top cover.
Rotate the observation light path selection knob b counterclockwise to remove it. (This selection knob should be
retained by the user because it will be reused when restoring the light path switching stroke to the original setting.)
Remove the hexagon socket head cap screws (4 x 16 ) d (4 positions) of the observation light path unit c using
the Allen wrench (3 mm), and place the unit c upside down.
Remove the screws f (2 positions) that clamp the filter frame e using the Phillips screwdriver and remove the filter
frame e .
Insert the filter g into the filter holding slit (long slit) of the filter frame.
The filter frame is slightly deformed toward outside, but this is normal.
Secure the filter frame e at the original position.
1
2
3
4
5
TIP
6
e
b
a
g
d
d
c
f