MX63/MX63L
109
The units described below must be
assembled and adjusted
by Olympus.
If these units are assembled or adjusted by the customer, the operations are not ensured.
Appendix: Units attached by Olympus
Remove the stage.
Insert the provided hexagon socket head cap screws (3 x 25) into four screw holes on the condenser a putting the
washer between, and insert the condenser downward into the opening b of the focusing unit so that it faces the
front (i.e. the AS lever or shutter comes to the front).
Place the condenser by pushing it toward left and back sides when looking at the microscope from the front, and
tighten the four hexagon socket head cap screws (3 x 25) using the Allen wrench (2.5 mm). The attaching of the A
type is completed.
For the B type, secure the interlock section c of the condenser height adjustment ring temporarily by inserting the provided
hexagon socket head cap screws (3 x 5) into four screw holes on the front of the focusing unit using Allen wrench (2.5 mm).
Then, secure the interlock section at the position where the gear interlocks smoothly.
Loosen the hexagon socket head cap screws (3 x 6) of the cover plate on the right of the base using the Allen wrench
(2.5 mm), and remove the cover plate.
Raise the focusing unit to the upper limit, align the dovetail groove e of the illuminator unit d with the internal
dovetail, insert it until it touches the end, and tighten the clamping screws of the internal dovetail using the Allen
screwdriver. In addition, to prevent loosening the condenser, screw in the provided hexagon socket head flat point set
screws (5 x 6) over the clamping screws. (using Allen wrench (2.5mm))
Secure the illuminator unit cover f using the hexagon socket head cap screws (3 x 6) provided with the microscope
frame so that the spaces of the openings are uniform with respect to the protruded part of the illuminator unit d .
Loosen the clamping screw g of the light guide holder, and pull out the light guide holder by holding the knob h .
Insert the light guide i and secure the clamping screws on the side.
Attach the light guide holder at the original position and secure the clamping screw g .
1
2
3
4
5
6
7
8
9
10
Units must be attached carefully to prevent them from
leaving fingerprints or scratches.
NOTE
c
a
b
d
e
f
i
g
h
1
Transmitted light illumination unit (MX-TILLA/MX-TILLB)