DMTA-10055-01EN, Rev. A, February 2015
Software Features and Options
273
Run mode is also useful when the sound path from the initial pulse to the reflection in
the interface echo varies because of transducer or test piece movement or other
factors. In run mode, any variation in the position of the echo inside the interface gate
is not visible, as this echo is actively held to the left side of the display. This allows
more stable visibility to the test material reflection positions regardless of interface
variation.
Note that the indication being tracked by the interface gate is determined by the gate
measurement mode. For instance, if the interface gate is in edge detection mode, the
tracked indication is the first echo to break the interface gate threshold.
To adjust the interface gate run mode
1.
Navigate to the
GateIF Setup
menu item (group 2/5).
2.
Press
GIF Run
(P3 key).
3.
Use the adjustment knob or arrow keys to set the parameter to
On
(see
Figure 11-34 Setting the gate interface run mode
Summary of Contents for EPOCH 650
Page 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Page 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Page 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Page 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Page 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Page 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Page 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Page 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Page 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Page 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Page 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Page 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...