DMTA-10055-01EN, Rev. A, February 2015
Calibration
171
5.
Once a steady reading is achieved, choose
Auto Cal > Cal-Zero
.
The screen freezes and the
Cal-Zero
box appears.
Figure 9-21 Entering the Zero Cal thickness value
6.
Adjust the value to match the known thickness of the gated indication (4.000 in. in
this example), and then choose
Continue
to continue to the second calibration
step.
If for any reason you need to exit without acquiring calibration data, press
Cancel
.
7.
Use the G1Start button (P5 key) to position gate 1 so that the second echo
reflection from the arc of the block is within the gated region. This reflection
should be close to 225 mm (9 in.) (see Figure 9-22 on page 172).