DMTA-10055-01EN, Rev. A, February 2015
Chapter 11
270
Initially, this base gain level is activated to show the same gain level as the primary
gain control. If the primary gain control consists of a reference gain setting and
scanning dB, the base gain level of the BEA will be the sum of the reference and
scanning gain levels. You can adjust the BEA gain independently to suppress the back
wall echo. Measurements and alarms under gate 2 are based on the attenuated echo.
To adjust back wall gain
1.
Press
BEA Gain
(P6 key).
2.
Manually adjust the gain or press 2
ND
F, (AUTO XX%) to automatically adjust the
echo in the BEA gate to the reference height (XX%) (see Figure 11-32 on page 270).
3.
Use gate 2
Start
(P2 key),
Width
(P3 key), and
Level
(P4 key), to adjust the BEA
gate position.
4.
Use Zoom to expand the signal display while adjusting parameters (see
“Zooming on a Gate” on page 132.
Figure 11-32 Adjusting back wall gain
When BEA is active, Gate 2 can be positioned anywhere on the visible screen range, to
the extent of overlapping Gate 1.
Summary of Contents for EPOCH 650
Page 12: ...DMTA 10055 01EN Rev A February 2015 List of Abbreviations xii...
Page 30: ...DMTA 10055 01EN Rev A February 2015 Introduction 18...
Page 54: ...DMTA 10055 01EN Rev A February 2015 Chapter 1 42...
Page 84: ...DMTA 10055 01EN Rev A February 2015 Chapter 3 72...
Page 126: ...DMTA 10055 01EN Rev A February 2015 Chapter 5 114...
Page 148: ...DMTA 10055 01EN Rev A February 2015 Chapter 7 136...
Page 154: ...DMTA 10055 01EN Rev A February 2015 Chapter 8 142...
Page 236: ...DMTA 10055 01EN Rev A February 2015 Chapter 10 224...
Page 332: ...DMTA 10055 01EN Rev A February 2015 Appendix B 320...
Page 342: ...DMTA 10055 01EN Rev A February 2015 Appendix C 330...
Page 346: ...DMTA 10055 01EN Rev A February 2015 Appendix D 334...
Page 362: ...DMTA 10055 01EN Rev A February 2015 Index 350...