DMTA-10055-01EN, Rev. A, February 2015
Chapter 9
148
Figure 9-2 Entering the Zero Cal thickness value
8.
Adjust the value to match the known thickness of the gated indication (0.200 in. in
this example), and then press
Continue
(P3 key) to continue to the second
calibration step (see Figure 9-3 on page 149). The thickness value used in this part
of the calibration will be stored for reference in the
Cal-Zero
parameter box.
To exit without acquiring calibration data, press
Cancel
(P1 key).