DMTA-10055-01EN, Rev. A, February 2015
Calibration
153
Figure 9-6 Example of a gated signal for zero calibration
Ensure that it is the first back-wall echo being gated, not a multiple echo from the end
of the delay line tip.
10. Once a steady reading is achieved, choose
Auto cal > Cal-Zero
.
The screen freezes and the
Cal-Zero
box appears (see Figure 9-2 on page 148).